TADT-1G Portable Humidity Generator(5L)
Introduction
Key Features
· The test chamber offers a volume 3 to 5 times larger than comparable products, enabling more efficient and flexible calibration conditions.
· It is compatible with a wide range of immersion-type chilled mirror dew point sensor heads, allowing for convenient comparative calibrations.
· Modular sensor ports accommodate temperature and humidity probes and transmitters of various diameters, enhancing versatility.
· A transparent observation window combined with an integrated lighting system facilitates easy reading of small-scale temperature and humidity instruments.
· Supports simultaneous calibration of 10 to 15 miniature temperature and humidity probes, or 3 to 5 standard-sized digital temperature and humidity meters.
· Ensures temperature uniformity of ≤ 0.1°C across the full calibration range of 5°C to 50°C.
· The externally mounted drying cylinder allows for desiccant replacement during operation, improving ease of maintenance and operational efficiency.
· Dual-zone drying cylinder design supports both molecular sieve and silica gel desiccants for flexible drying configurations.
· Real-time display of liquid level and desiccant status enables intuitive monitoring and timely maintenance.
· Equipped with four external signal input channels, supporting standard signal types: 0 - 1 V, 0 - 5 V, and 4 - 20 mA.
· Built-in humidity conversion software simplifies conversion between relative humidity, temperature, and dew point temperature.
· Temperature and humidity curves are displayed for real-time trends monitoring.
· Features a 7-inch full-color LCD touchscreen for clear visualization and user-friendly operation.
Technical Parameters
|
Model |
TADT-1G |
TADT-2G |
TADT-3G |
|
Humi range |
5%RH ~ 95%RH(20±3℃) |
||
|
Temp range |
5℃ ~ 50℃ |
||
|
Cavity working size(mm) |
160×175×175 |
280×175×175 |
110×110×110 |
|
Volume |
5L |
9L |
1.5L |
|
Overall dimensions(mm) |
550×300×600 |
550×300×600 |
370×480×200 |
|
Max allowable error of temp. |
±0.05℃(20±3℃) ±0.1℃(5~50℃) |
||
|
Max allowable error of RH |
±0.8%RH(20±3℃) ±1.5%RH(5~50℃) |
||
|
Temp. uniformity |
≤ 0.1℃(20±3℃) |
||
|
Humi .uniformity |
≤ 0.5%RH(20±3℃) |
||
|
Temp. stability |
±0.05℃ |
||
|
Humi. stability |
±0.5%RH |
||
|
Max number of openings in the test window |
13 pcs (standard hole φ15mm, quantity 12, φ48mm quantity 1pcs) |
||
|